Laboratory Quality & Accreditation Readiness

Prepare for NABL accreditation by proving technical competence for each test or calibration in the proposed ISO/IEC 17025 scope. Trace a released result to its method, authorised person, equipment status, original observations, quality controls and applicable decision rule.

  1. Define the scopeMethod, parameter, matrix or instrument, range and location
  2. Trace a resultSample, operator, equipment, original data and calculations
  3. Review the statementUncertainty, customer agreement, decision rule and authorisation

Implementation guidance

How do you build an ISO/IEC 17025 scope matrix?

For each proposed entry, capture the method and version, measurable parameter, product or matrix, applicable range or capability, and responsible technical staff. Tie these to verification or validation evidence, equipment and traceability, and the planned external comparison or proficiency testing (PT). Check the latest relevant NABL documents for the type of laboratory and scheme. Start with work the laboratory can demonstrate consistently.

What is the difference between method verification and validation?

Verification checks whether the laboratory can achieve the required performance when it correctly implements an established method. Validation considers whether the method is fit for the intended use in cases such as a laboratory-developed or modified method. Set the acceptance criteria before collecting evidence. Check for changes to sample matrix, range, software, equipment or personnel that could affect performance, and decide what to review again.

How does a decision rule affect a conformity statement?

A decision rule sets how measurement uncertainty is considered when reporting whether a result conforms to a specification. Document the rule and agree it with the customer when required, unless the relevant specification or standard already prescribes it. Confirm the limit, direction of conformity, units and scope of the statement. In technical review, ask whether another authorised reviewer can reach the same reporting decision from the same results.

Worked example

A result near the tolerance limit

An illustrative dimensional test has an upper limit of 10.00 mm. The measured result is 9.98 mm with an expanded uncertainty of 0.05 mm. The upper end of that interval is 10.03 mm. If the agreed guard-band rule accepts only when measured value plus expanded uncertainty is at or below the limit, this result fails to meet the acceptance criterion. The report should follow the agreed rule, including how it describes borderline or indeterminate outcomes.

Trace a laboratory result back to its raw observations

An ISO/IEC 17025 readiness exercise can start with a single issued result and trace the sample ID, method version, instrument, raw data, calculation and authorised review. Ask a second reviewer to reproduce the reported value from the retained records.

  1. Select a result with a calculation

    Record the report number and sample ID. Retrieve the raw file and approved calculation worksheet or software configuration used on the day the result was generated.

  2. Check qualification and instrument status

    Review the analyst's authorisation for the method at the test date, along with the instrument status and relevant quality-control results for that run.

  3. Review amendments

    If the final number was revised, verify who authorised the change, why it was needed and how the current issue is linked to the earlier issue. Retain the record trail for technical review.

Evidence checklist

FocusWhat to checkEvidence or decision
ScopeMethod, matrix or range and locationApproved scope entry and method evidence
Technical resultRaw readings, equipment and calculationsTraceable certificate or record, control results, reviewer check
Conformity statementApplicable requirement and decision ruleCustomer agreement when needed and report sign-off

Common questions

Is every result from a NABL-accredited laboratory an accredited result?

Accreditation applies to the published scope. Check the test or calibration, method, location and relevant range or capability against that scope.

Is one passing proficiency test enough to show ongoing competence?

Treat it as one evidence point. Review relevance, PT coverage, internal quality-control trends and corrective action where performance was unsatisfactory.

How can a lab check the integrity of a spreadsheet calculation?

Recalculate a representative result independently using approved inputs, units and rounding. Record which worksheet version was used and who reviewed the check.